About Us Contact Us
 
 
   
OT-Surgery Equipments - Anesthesia Delivery System
 
Asteros Lite 9000
 
Features :
1. 3 gas based-O2, N2O & Air with dual cascaded 5 tube rotameter with back light.
2. Designed for low flow anesthesia delivery.
3. Anti hypoxic guard for controlling gas ratio.
4. Cylinder and pipeline gas connectivity for O2 & N2O & air.
5. Large diameter high pressure gauge - 1each for O2, N2O & Air.
6. 2nd Stage pressure regulation to ensure absolute accuracy of gas delivery.
7. Autoclavable twin jar CO2 absorber (optional).
8. Provision for integrated anesthesia ventilator (optional)
9. Selected compatible vaporizer.

 

PHYSICAL:

Construction: Tabular, Rigid Thermostatmoulded with large drawer for storage, Four Large diameter anti-static castor wheels providing easy mobility. Front Castors have brakes.

GAS INLETS

  Cylinder yokes: Gas specific , pin indexed yokes with slinding steel clamping bar for O2 and N2O as per model below:Asteros Lite 9000: 2 X O2 ,2 X N2O . Central Pipeline Connections for Oxygen, Nitros Oxide and Air.   

REGULATORS:

Highly efficient and reliable regulators (2 stage in for O2 and N2O. N2O regulators are activated only if O2 regulators are pressurized. N2O cut Off mechanism: N2O regulators cut off on O2 supply failure

PRESSURE GAUGE CYLINDER :

O2 and N20 : 0 to 250 kg/cm2

PRESSURE GAUGE:

O2/N2O:0-10kg/cm2 

O2/N2O piping station block with pressure gauge and 5m hose (Standard). 

ROTAMETER

5 Tube long (230 mm) rotating bobbin flowmeters calibrated in double scale for accuracy and clarity in reading.

3 gas flowmeters O2, N2O, Air.

Anti-hypoxia system: Minimum 25% O2 at about 3kg/cm2.

Scale: O2 : 0.1-10LPM  N2O : 0.1-10LPM  

N2O SAFETY CUT-OFFVALVE FOR OXYGEN SUPPLYING FAILURE

When the pressure of the supplying oxygen drops bellow about 2.5kg/cm2, nitrous oxide will be automatically shut off to protect a patient from the danger of hypo-oxygenation

OXYGEN FLUSH VALVE

Self-returning push Button  Flow : 35-75 LPM at 4 kg/cm2.

ALARM WHISTLE FOR OXYGEN SUPPLYING FAILURE

Whistle more than 7 seconds to warn a danger when the pressure of supplying oxygen drops bellow about 2.5kg/cm3.

MOBILITY:

Four large diameter anti-static castor wheels with ball bearings. Front castors have brakes actuated by horizontal bar doubling as foot rest.

VAPORIZER

 Goldman Halothane vaporizer supplied as standard. (Standard with Asteros Lite 1100.

OPTIONAL ITEMS: 

1) Gas Attachment: Mox regulator with High Pressure Tubing for O2

 2) Circle Absorber: International Standards circle system with highest absorption efficiency. Easy to refill & change canister. Using design allows for use of single or dual canisters. Inbuilt APL & Pressure gauge with WYE/Extended Bains Breathing circuit.

3)Anesthesia Ventilator Adult/Pediatric Compatible.

 

 
Authorised Sales & Service Dealer
  • Works with Ultrasound
  • C-Arm
  • Lithiotripter
  • OPG
  • CR Systems
  • ICU Monitoring Systems
  • X-Ray Machines
  • Surgical Instruments
  • Darkroom
  • Accessories
  • High Voltage Transformers
  • X-ray Machine & Spares
 
Call +91-9810024291
 