Patient Range
Adult, Pediatric and neonatal
Performance Specification
Display : 12.1” color active matrix TFT
Resolution : 800x600
Trace : 7waveforms
Sweep Speed :12.5,25,50mm/s
Alarm indicator light
Power indicator light
Audio Indicators for QRS beep and alarm sound
Interface : Networking
Battery : Rechargeable
Trend time : 1~72 hours
Alarm : 3-level audible and visual alarm
Recorder (optional) : Built-in, thermal array,3channels
Record Width : 40mm (record paper 50mm)
Record Speed : 25mm/s, 50mm/s
ECG
Lead Type : 5-lead
Input : 5-lead(RA; LA; RL; LL; V)
Lead Selection : 5-Lead; I; II; III
ECG Waveform : 1 channel
Gain Selection : x0.5,x1,x2,auto
Sweep Speed : 12.5mm/, 25mm/s
Heart rate range : 25~20BPM
Accuracy : 1BPM
Anti-electrosurgical interference and defibrillation
Standard Configuration
ECG, RESP,NIBP,TEMP,SPO2
EtCO2 Micro Stream Latest Technology Modular Cassete.
OPTIONS:
IBP,FHR Module, Thermal, Recorder Li-ion Battery, Wall Mounting, Trolley
CMRR : Diagnostic mode:>60db
Monitor mode:>60db
S-T detection : YES
Measurement range : -20mV-2.0mV
Arrhymia analysis :YES
Alarm :YES, audible and visual alarm, alarm events
recallable
Respiration
Method : RA-LL impedance
Measurement Range : 20~250BrPM
Resolution : 1BrPM
Accuracy : 2% or 2BrPM, whichever is greater
Apnea Alarm : YES, and apnea delay:10~40seconds
NIBP
Method : Oscillometric
Operation Modes : Manual /Automatic
Measurement Unit : mmHg/kPa selectable
Measurement Type : Systolic pressure Diastolic pressure and Mean Pressure
Measurement Range : Systolic Pressure:50-24 mmHg
Diastolic Pressure:25~180mmHg
Mean Pressure:30~200mmHg
Over-pressure Protection : YES
Resolution : 1mmHg
Alarm : Systolic, Diastolic and Mean
Temperature
Scale : C and F Selectable
Measurement Range : 27’C ~45C
Resolution : 0.1 or
Channel : 1 Channel
SPO2
Range : 0~100%
Accuracy : 70% ~100%(+2%)
0%~69% : unspecified
Pulse Rate
Range : 20~254BPM
Accuracy : 3 BPM
Safety
Meet requirement of IEC60601-1
Dimension and weight
Dimension 335x270x200(mm)
Weight 5.4 kg
Power Supply
Power Source : AC mains power or Internal battery power
Power Requirements : AC 220V(10%)
Line Frequency : 50 Hz (+ 1Hz)
Consumption : 60 W
Battery Power
The maximum number of installed battery:1
Operating time : 180 minutes under the normal use and full
charge
Operation Environment
Temperature :10C to 30’C (50F to 86F)
Humidity :15% to 70%, non-condensing