About Us Contact Us
 
 
   
Critical Care - Ventilator - Portable Ventilator
 
Falco 202
 
Features :
 

 

Application Equipment for ventilation. It can also be used during

transportation or in hospital departments.
Patients Adults / Children
ADJUSTABLE MODES AND PARAMETERS
Operation Modes CPAP, VAC, SIMV, BILEVEL-S, BILEVEL-ST, PCV, (A)PCV, PSV, PSV-TV
with PSV and PS (Tv) modalità with guaranteed rate and also for PS(Tv)
with a guaranteed volume.
IPAP - PS 6 - 60 (cmH2O, hPa)
EPAP - PEEP 0 - 20 (cmH2O, hPa)
Rate 5 - 50 (RPM)
Inspiratory Time 0,4 - 3 (sec)
Min. insp. time 0,3 - 2,5 (sec)
Max. Insp time 1 - 3 (sec)
Insp. Ramp Slope 0, 1, 2, 3 (ACCELERATION SLOPE)
(0 max acceleration)
Sweet Start Option 0-50 (min) (5 minute steps) it increases gradually the pressure to reach the
setting value
Inspiratory Trigger
(flow trigger)
1 – 6 (l/min)
Expiratory Trigger
(flow trigger)
5 – 90 (%) (relative to maximum value of inspired flow)
Safety Tidal Volume 50 - 2500 (ml)
Falco 202_te_rev0 2 / 4
Tidal volume 50-2500 (ml)
MESURED PARAMETRES
Airway pressure IPAP (Inspiratory Positive Airways Pressure), EPAP (Expiratory Positive
Airways Pressure) (–20 80) (cmH2O, hPa)
Rate 0 – 99 (RPM)
I:E Ratio 1:4 – 4:1
Exp. Tidal Volume 0 - 9000 (ml)
Exp. Minute Volume 0 – 40 (l/min)
PARAMETERS STORAGE
TRENDS Storage capacity (72 h) of all mesured parameters.
EVENTS Storage capacity (72 h) of all machine’s events and all alarms.
ALARMS
Airway Pressure High, Low
Rate High, Low
Inspired Tidal Volume High, Low
SYSTEM ALARMS
Turbine overtemperature Signals in case of high turbine temperature
Turbine fault Signals in case of a blower fault condition
Apnoea
Alarm condition: RATE=0. In case of Apnoea, in PSV, PS(Tv), Bilevel S
modes, the ventilator give a controlled ventilation (APNOEA BACKUP). This
alarm is present only in Bilevel S mode. In the others assisted breaths are
present; a min. guaranteed rate setting by RRbk parameter.
Power supply fault Alarms in case of power supply fault
Charge battery (NiMh) level Battery (NiMh) status in steps of 25% charge
Battery “LOW” Alarms, with a non - resettable alarm , the condition in which the remaining
time charge of the battery is < 10 minutes.
Battery overtemperature Alarms in case of high battery temperature
Battery disconnect Cannot reset alarm condition: battery fail
Machine Hours Maintenace Low priority alarm: 2000 work hours
INITIAL TESTS
Turbine Test The correct functioning of the turbine is tested
Flow sensor test The correct functioning of the flow sensor is tested
Pressure sensor test The correct functioning of the pressare sensor is tested
Recharging battery level
test
The level of battery charge is tested
Electrovalve test The correct functioning of electrovalve is tested
Leakage test / Patient
circuit
Presence of Leakages on patient circuit and inside the unit is tested
Falco 202_te_rev0 3 / 4
Fan test The correct functioning of the cooling fan is tested
Audio alarm test The correct functioning oth the audio buzzer is tested
CONNECTIONS – POWER SUPPLY
Patient circuit Single tube or Double tube patient circuit with expiratory valve; sterilizable
or not sterilizable.
Power Supply 12 VDC / 100-240 VAC; Imax= 7,6A; Pmax= 105 W
Built-in Battery 12 VDC NiMh 1 pack 4.5 Ah (autonomy: 2.30) (optional)
Additional Battery 12 VDC external NiMh (autonomy: 10 h) (optional)
External connections N.1 RS232 output, one output for external alarm, n.1 RJ45 interface to
INTERNET through TCP/IP protocol.
Expiratory flow sensor
connection N.1 input to connect the expiratory flow sensor at the ventilator.
PHYSICAL CHARACTERISTICS
Noise 35 dB
Dimensions (L x P x H) 240 x 330 x 210 mm
Weight 3,9 Kg
Environmental conditions Temperature from -10 to 40 °C
Non condensing relative Humidity from 15 to 95 %
Atmospehric pressure from 70 to 110 Kpa
Conformity to Norms EN 1281-1, IEC 601-1, IEC 601-1-2, EEC 93/42 Dir.ve, IEC 601-1-4,
EN 794-2
Class and type according to
IEC 601-1
Class 1 Type B
Class according to 93/42
EEC Dir.
Class IIb
Protection IP degree IP 44
STANDARD ACCESSORIES
Manual Operator user manual + Patient user manual
Patient circuit Coaxial PVC patient circuit cm 180 (n.1 pc supplied with the unit)
Aspiration filtre Turbine antipollution aspiration filtre (n.1 pc supplied with the unit)
Aspiration filtre Turbine antidust aspiration filtre (n.1 pc supplied with the unit)
Patient antibacterial filtre Antibacterial filtre for patient circuit (n.1 pc supplied with the unit)
Fan filter Air fan cooling filter (n. 1 pc supplied with the unit)
Power supply Power Supply 13,8V- 7,6A 105W with connection cable
Vehicular cable 12 VDC cable with lighter connector
Auxiliary Battery NiMh Battery 4,6 Ah – 2,5h autonomy (optional)
Flow sensor Expiratory flow sensor
OPTIONAL ACCESSORIES
Falco 202_te_rev0 4 / 4
Patient circuit Single tube sterilizable cm 180
Patient Circuit Double hose PVC patient circuit adult cm 180x22 +Y
Patient circuit Coaxial PVC patient circuit cm 180
Expiratory valve for single
tube patient circuit Expiratory valve (with PAW inlet)
Expiratory valve for double
tube or coaxial patient
circuit
Expiratory valve
Auxiliary Battery NiMh Battery 4,6 Ah – 2,5h autonomy
Aspiration filtre Turbine antipollution aspiration filtre (n.1 pc supplied with the unit) (pack 10
pc.)
Aspiration filtre Turbine antidust aspiration filtre (pack 10 pc.)
Patient antibacterial filtre Antibacterial filtre for patient circuit (n.1 pc supplied with the unit)
Humidifier VAPOUR 2 electronic humidifier
Fan filter Air fan cooling filter (n. 1 pc supplied with the unit)
Power supply Power Supply 13,8V- 7,6A 105W with connection cable
Vehicular cable 12 VDC cable with lighter connector
Transport bag Transport bag for Falco 202
Ethernet connection Connection module for Ethernet board
Auxiliary Battery NiMh Battery 4,6 Ah – 2,5h autonomy (n.1 package supplied with the unit)
Flow sensor Expiratory flow sensor
Trolley CASTOR LT trolley with housing for the ventilator and its accessories. Rail
for humidifier fixing.
 
 
Authorised Sales & Service Dealer
  • Works with Ultrasound
  • C-Arm
  • Lithiotripter
  • OPG
  • CR Systems
  • ICU Monitoring Systems
  • X-Ray Machines
  • Surgical Instruments
  • Darkroom
  • Accessories
  • High Voltage Transformers
  • X-ray Machine & Spares
 
Call +91-9810024291
 
Copyright © 2010 AA Medical Systems. All Rights Reserved.